Home > Products > OTO-Chek Biological Test Simulator

OTO-Chek Biological Test Simulator

Company : AMBCO Electronics

UMDC Code : 102475

Simply place the earphones over right and left couplers of OTO-CHEK. , For MANUAL TESTING turn OTO-CHEK on. Present the tones as in manual testing and record the frequency and HL when there is positive RESPONSE by OTO-CHEK, red light on. Retain these baseline results and use them to compare future tests. ESTABLISH NEW BASELINE every time the audiometer is re-calibrated. For AUTOMATIC TESTING place the earphones on OTO-CHEK, plug the response switch from OTO-CHEK to response jack on audiometer. Turn OTO-CHEK on. Turn automatic audiometer on and initiate automatic test.If printer is connected to the audiometer, results will be printed. Retain these baseline results and use them to compare future tests. ESTABLISH NEW BASELINE every time the audiometer is re-calibrated.

Product URL : http://www.ambco.com/audiometers/OTO.html

Last Updated : 03/16/2013

Please provide the following information and a representative of AMBCO Electronics will contact you to assist with any questions you might have.