Home > Products > 270-T · Model 270 · Type T · -200 to 250C

270-T · Model 270 · Type T · -200 to 250C

Company : J-KEM Scientific Inc

UMDC Code : 357389

If the reaction reaches the high temperature limit or a thermocouple breaks, power is turned off to the heater until the controller is manually reset. Both meters independently monitor the reaction temperature. In the event that one meter fails, the other takes over to prevent a heating accident. J-KEMs design provides independent, 100% redundant control.Designed for processes requiring uncompromising safety. The Model 270 has a built-in, independent, backup controller to guard against heating accidents from equipment failure. The desired temperature is entered into the main temperature controller which regulates heating, and then a high temperature cut-off is entered into the backup limit controller.

Product URL : http://www.jkem.com/lab-safety-equipment/safety-te

Last Updated : 03/16/2013

Please provide the following information and a representative of J-KEM Scientific Inc will contact you to assist with any questions you might have.