Alkaline Phosphatase (ALP). Alkaline phosphatase is widely distributed in the body and is present in high concentrations in bone, intestinal mucosa, and renal tubule cells. Less high concentrations appear in the liver, leukocytes, and placenta. Increased values for alkaline phosphatase are found in all bone disorders, liver disease, and during the third trimester of pregnancy. Decreased values are found in hypophosphatasemia, hypothyroidism, pernicious anemia, and in dwarfs.
|
Company:
Date:
November 12, 2008
|
|
|
|
|
|
|
|
|
|
PRINCIPLE
The enzymatic sequence employed in the assay of Alkaline
Phosphatase is as follows:
Alk. Phos.
p-Npp H20 --------------- p-Nitrophenol H3PO4
p-Npp is colorless but p-Nitrophenol has strong absorbance at 405 nm.
The rate of increased absorbance at 405 nm is proportional to the
enzyme activity.
|
Company:
Date:
November 12, 2008
|
|
|
|
|
|
|
|
|
|
PERFORMANCE CHARACTERISTICS
Analytical Range: 5-1000 IU/L
For alkaline phosphatase analyte using the Teco Alkaline Phosphatase Reagent on SYNCHRON CX® System, this method has been demonstrated to be linear from 5-1000 IU/L
Accuracy: Comparison study was performed on Beckman CX® System using 40 samples. Beckman Coulter Alkaline Phosphatase reagent was used to compare with Teco Alkaline Phosphatase Reagent. The results of this study in yield a correlation coefficient of 0.99 with a regression equation of y=1.1X -2.6.
Precision:
Within Run precision for Teco Alkaline Phosphatase Reagent Set was d
|
Company:
Date:
November 12, 2008
|
|
|
|
|
|
|
|
|
|
PERFORMANCE CHARACTERISTICS
Analytical Range: 0-900 IU/L
For Alkaline Phosphatase analyte by Alkaline Phosphatase Reagent on
TC Matrix, this method has been demonstrated to be linear from
5-1000IU/L
Accuracy: Comparison study was performed on TC Matrix from 40
samples. Beckman Coulter Alkaline Phosphatase reagent was used to
compare with Teco Alkaline Phosphatase Reagent. The results of this
study in yield a correlation coefficient of 0.99 with a regression
equation of y=0.99X 1.9.
Precision: Within Run precision for Alkaline Phosphatase Reagent Set
was determined following a modification of N
|
Company:
Date:
November 12, 2008
|
|
|
|
|
|
|
|
|
|
Alkaline Phosphatase (ALP)
|